RF emissions
CISPR11/EN55011
Group 1
RF emissions
CISPR11/EN55011
Class B
Harmonic emissions
IEC61000-3-2/EN61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC61000-3-3/EN61000-3-3
Complies
Guidance and manufacturer's declaration - Electromagnetic Emissions.
Emissions test Compliance Electromagnetic environment - guidance
Surgic Pro series is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
ޓ
Surgic Pro series uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
Surgic Pro series is intended for use in the electromagnetic environment specified below. The customer or the user of the Surgic Pro series should assure that is
used in such an environment. Guidance and manufacture’ sdeclaration electromagnetic emissions.
Surgic Pro series is intended for use in the electromagnetic environment specified below.
The customer or the user of the Surgic Pro series should assure that it is used in such an environment.
Electrostatic discharge (ESD)
IEC61000-4-2/EN61000-4-2
± (2, 4) 6 kV contact
± (2, 4) 8 kV air
± (2, 4) 6 kV contact
± (2, 4) 8 kV air
Electrical fast transient/burst
IEC61000-4-4/EN61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
± 2kV for power supply lines
± 1kV for input/output lines
Surge
IEC61000-4-5/EN61000-4-5
± 1kV differential mode
± 2kV common mode
± 1kV differential mode
± 2kV common mode
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11/EN61000-4-11
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
Power frequency
(50/60Hz) magnetic Field
IEC61000-4-8/EN61000-4-8
3A/m 3A/m
Guidance and manufacturer's declaration - Electromagnetic Immunity
NOTE ‘Ut’ is the AC mains voltage prior to application of the test level.
Immunity test IEC/EN60601 test level Compliance level Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the Surgic Pro series
requires continued operation during power mains
interruptions, it is recommended that the Surgic Pro series
be powered from an uninterruptible power supply or a
battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
15. Symbols
16. EMC Information (Electromagnetic Compatibility Information)
TUV Rhineland of North America is a Nationally Recognized Testing Laboratory (NRTL) in the United States and is accredited by the
Standards Council of Canada to certify electro-medical products with Canadian National Standards.
This product is designed not to become the ignition source in air and flammable anesthetic gas.
The EU directive 93/42/EEC was applied in the design and production of this medical device.
Protected against the effects of continuous immersion in dust and water.
Dispose of this device and its accessories via methods approved for electronic device and in compliance with the Directive 2002/96/EC
Marking on the outside of Equipment or Equipment parts that include RF transmitters or that apply RF electromagnetic energy for
diagnosis or treatment.
Type BF applied part
See Operation Manual
English
21