4-4 Connecting the AC Power Cord
Ensure the Pump Cover Lever is in the ‘OPEN’ position, with the Irrigation tube ‘Needle’ towards the backside of the
Control Unit. Position the stoppers of the tube assembly in to the Tube Guide securely. (Fig.4)
4-5 Installing the Irrigation Tube
Correctly align then insert the AC Power Cord into the
AC Power Cord Connector Socket at the back of the
Control Unit. (Fig.3)
4-2 Connecting the Handpiece
Push the handpiece on to the motor taking care to keep it in
alignment.
Twist the handpiece until it locks into position with a clicking sound.
1)
2)
Face the screw on the Foot Control Cord Plug downward
then insert the plug into the Foot Control Cord connector
socket on the Control Unit. Secure the plug by fastening
the Lock Nut. Refer Fig.3.
4-3 Connecting the Foot Control
CAUTION
<OPEN Position>
AC Power Cord
Foot Control Cord Plug
Screw
Water flow direction
Pump Cover
Tube Guide
Fixed Position
Fixed Position
Fig.2
Fig.3
Fig.4
When using the Optic Surgic Pro micromotor (SGL70M), USE ONLY Optic contra-angle handpieces. Non Optic
handpieces may not precisely connect to the micromotor and it may malfunction when the micromotor is under
load.
Stopper
Pump Cover Lever