Page #39 background imageLoading...
Page #39 background image
38
22
Disposing product
In order to avoid the health risks of operators handling the disposal of medical equipment, as well as the risks of environmental contamination
caused thereof, a surgeon or a dentist is required to confirm the equipment is sterile. Ask specialist firms who are licensed to dispose of
specially controlled industrial wastes, to dispose the product for you.
23
EMC Information (Electromagnetic Compatibility Information)
Guidance and manufacturer’s declaration - Electromagnetic Emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure
that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR11/EN55011
Group 1 The product uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR11/EN55011
Class B The product is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
EN/IEC61000-3-2
class A (except 120V)
Voltage fluctuations/flicker
emissions
EN/IEC61000-3-3
Complies (except 120V)
Guidance and manufacturer’s declaration - Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure
that it is used in such an environment.
Immunity test EN/IEC60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
EN/IEC61000-4-2
±(2,4)6kV contact
±(2,4)8kV air
±(2,4)6kV contact
±(2,4)8kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast transient/burst
EN/IEC61000-4-4
±2kV for power supply lines
±1kV for input/output lines
±2kV for power supply lines
±1kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
EN/IEC61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN/IEC61000-4-11
<5% Ut
(>95% dip in Ut)
for 0.5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 0.5 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
<5% Ut
(>95% dip in Ut)
for 0.5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 0.5 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the product requires continued
operation during power mains interruptions, it
is recommended that the product be powered
from an uninterruptible power supply or a
battery.
Power frequency
(50/60Hz) magnetic field
EN/IEC61000-4-8
3A/m 3A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: ‘Ut’ is the AC mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration - Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure
that it is used in such an environment.
Immunity test EN/IEC60601 test level Compliance level Electromagnetic environment - guidance

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NSK Varios Combi Pro and is the answer not in the manual?

NSK Varios Combi Pro Specifications

General IconGeneral
BrandNSK
ModelVarios Combi Pro
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals