English
35
15
Specifications
<Control Unit>
Model Varios Combi Pro
Power source AC100V - 240V 50/60Hz
Drive frequency 28 - 32kHz
Power input 78 - 98VA
Maximum output 11W
Supply air pressure 0.55 - 0.75MPa (5.5 - 7.5kgf/cm
2
)
Supply water pressure 0.1 - 0.3MPa (1.0 - 3.0kgf/cm
2
)
Bottle volume 400ml
Irrigation Volume (Ultrasonic Mode) 5.0 - 70.0mL/min (Irrigation Mode: Bottle/Tap water, Supply water pressure: 0.2MPa)
Dimension W225 × D290 × H162mm (Including the Irrigation Bottle and Chamber)
Weight 2.7kg (Irrigation Bottle and Chamber not included)
<Foot Control>
Model FC-79
Cord length 2.5m
Dimensions W204 × D157 × H128mm (including Hanger)
Weight 415g (including Hanger)
<Ultrasonic Handpiece>
Model VA2-LUX-HP
Cord length 1.8m (Connector not included)
Type of oscillation Piezo type
Light White LED
Dimensions
20 × 103mm (cord not included)
Weight 54g (cord not included)
<Powder Handpiece>
Model PROPHY-HP/PERIO-HP
Hose length 1.8m (Connector not included)
Dimensions
16 × 107mm (cord not included)
Weight 33g (cord not included)
Temperature Humidity Atmospheric pressure
Use environment*
2
0 - 40°C 35 - 75%*
1
–
Transportation and Storage
environment
–10 - 50°C 10 - 85% 500 - 1,060hPa
*1 Without condensation
*2 The use of this product under conditions other than specified above may cause malfunction.
16
Classification of Equipment
tType of protection against electric shock:
– Class l equipment
tDegree of protection against electric shock:
Type B applied part: (Applied parts: Ultrasonic Tip, Ultrasonic Handpiece, Nozzle Tip, Powder Handpiece)
tMethod of sterilization or disinfection recommended by the manufacturer:
Autoclave sterilization
tDegree of protection against ingress of water as detailed in the current edition of IEC 60529:
– Foot Control: IPX1 (Protected against vertically falling water drops)
tDegree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide:
– Equipment NOT suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
tMode of operation:
– Continuous operation