8-2 Sterilization
Autoclave sterilization is recommended.
Autoclave sterilization is required for the first time use and after each patient as noted below.
The following items can be autoclaved.
Handpiece with Handpiece Cord (Optic/Non-Optic), Tip, Tip Wrench, Tips Holder, Handpiece stand and Tube Holder.
Note the following when sterilizing a diamond coated tip:
-
Follow your local directives, standards, and guidelines.
-
Sterilize a tip in a U.S. FDA-cleared sterilization pouch, compliant to ISO 11607-1.
-
NSK recommends sterilization according to ANSI/AAMI ST79 and EN 13060 (class B).
-
Follow the manufacturer's instructions for use that came with your sterilizer.
-
For more information about sterilization standards, refer to the medical device regulations specific to your location.
*
U.S.: The recommended sterilization procedures require the use of FDA-cleared sterilizers, sterilization trays,
sterilization wraps, biological indicators, chemical indicators, and other sterilization accessories labeled for the
sterilization cycle recommended
.
8-2-1 Handpiece
Optic Handpiece
1) Disconnect the Handpiece cord from the Control Unit.
2) Remove the Irrigation Tube from the Handpiece and dispose of it.
3) Brush (do not use a metal brush) dirt off the surface, and then wipe with alcohol soaked cloth.
Non-Optic Handpiece
1) Turn the Nose Cover counter-clockwise to remove.
2) Brush any contaminants from inside the Nose Area with a soft
brush. (Do not use metal brush) Wipe with alcohol soaked cloth
(Fig. 26).
3) Replace the Nose Cover onto the Handpiece turn clockwise.
CAUTION
Only handpiece with cord can be washed via Thermo Disinfector.
* Confirmed Thermo Disinfector is “Type: G7882” (Miele & Cie)
The Handpiece can not be removed from the
Handpiece Cord. Removing or twisting the Handpiece
could cause damage to the internal parts and failure
may occur.
CAUTION
Handpiece CordHandpiece
Nose Cover
Tighten
Loosen
Nose Area
Fig.26