20
Case Possible Cause Remedy
The Tip breaks easily. Power Level is not right for the Tip mounted.
Change to the correct mode and power level
as indicated on the Tip case.
The Tip comes off. The Tip is not tightened sufficiently Tighten the Tip until the Tip Wrench clicks.
Handpiece makes a loud noise.
Power Level is not right for the Tip mounted.
Change to the mode and power level
indicated on the Tip case.
Tip is not tightened sufficiently. Tighten the Tip until the Tip Wrench clicks.
Failure inside the Handpiece or the Control
Unit.
Contact dealer. *
Handpiece gets hot.
Power Level is not right for the Tip mounted.
Change to the mode and power level
indicated on the Tip case.
Tip is not tightened sufficiently. Tighten the Tip until the Tip Wrench clicks.
Failure inside the Handpiece or the Control
Unit.
Contact dealer.*
Foreign substance is clogging Handpiece
and blocking saline.
Place air syringe against Water Pipe and
blow air through it.
If this does not clear it, contact dealer.
Saline is not supplied.
Make sure bottle still has saline in it and
Irrigation Tube has no cracks.
Saline does not come out as
spray.
It depends on the combination of flow level,
power level, and Tip shape.
With some flow levels and Tip shapes, it is
more difficult to form a spray. This is not a
malfunction.
Water does not come out, or flow
level is low.
Foot Control is not connected. Make sure it is connected.
Irrigation Tube is not connected to bottle or
Handpiece.
Make sure they are connected.
Irrigation Tube is not properly mounted to
the pump.
Make sure it is mounted properly (Refer to
6-5-1 Mounting to the Irrigation Pump).
The Pump Cover is open. Close the cover.
The Irrigation Tube is broken (leaking).
Replace the Irrigation Tube (Refer to 6-5
Mounting the Irrigation Tube).
Water leak.
Leak from the connection between bottle
and Irrigation Tube.
Insert Irrigation Tube Needle all the way into
the bottle (Refer to 6-5-3 Mounting to the
Bottle).
Leak from the connection between
Handpiece and Irrigation Tube.
Insert end of Irrigation Tube all the way into
the Handpiece water pipe (Refer to 6-5-2
Mounting to the Handpiece).
Leak from Irrigation Tube.
Replace the Irrigation Tube (Refer to 6-5
Mounting the Irrigation Tube).
Irrigation doesn’t stop.
Mode is set to Auto Cleaning. To stop irrigation, press the Flow Key.
ON-OFF Switch for Water Flow on Foot
Control is being operated.
Remove foot from the Foot Control.
Faulty circuit. Contact dealer.*