23
Symbols
TUV Rheinland of North America is a Nationally Recognized Testing Laboratory (NRTL) in the United States and is accredited 
by the Standards Council of Canada to certify electro-medical products with Canadian National Standards.
The product is designed not to become the ignition source in air and flammable anesthetic gas.
*Only Foot Control is AP equipment.
The EU directive 93/42/EEC was applied in the 
design and production of this medical device.
Protected against the effects of continuous 
immersion in dust and water.
Dispose of this device and its accessories via methods approved for electronic device and in compliance with the Directive 
2002/96/CE.
Type BF applied part. Follow Operation Manual for use. Manufacturer
Authorised representative in the European community.
Autoclavable up to Max.135°C.
*for detail see Sterilization.
This product can be cleaned and disinfected with a Thermo-Disinfector.
Marking on the outside of Equipment or Equipment parts that include RF transmitters or that apply RF electromagnetic energy 
for diagnosis or treatment.
Caution: U.S.Federal law restricts this device to sale by or on the order of a licensed physician.
Guidance and manufacturer's declaration - electromagnetic emissions
The VarioSurg is intended for use in the electromagnetic environment specified below. The customer or the user of the VarioSurg should assure that is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR11/EN55011
Group 1 The VarioSurg uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment.
RF emissions
CISPR11/EN55011
class B The VarioSurg is suitable for use in all establishments other than domestic, and may be used in domestic establishments 
and those directly connected to public low-voltage power supply network that supplies buildings used for domestic 
purposes, provided the following warning is heeded:
Warning: VarioSurg is intended for use by healthcare professionals only. VarioSurg may cause radio interference or may 
disrupt the operation of nearby equipment. 
It may be necessary to take mitigation measures such as re-orienting or relocating the VarioSurg or shielding the location.
Harmonic emissions EN/IEC61000-3-2 class A
Voltage fluctuations/ flicker emissions
EN/IEC61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The VarioSurg is intended for use in the electromagnetic environment specified below. The customer or the user of the VarioSurg should assure that it is used in such an environment.
Immunity test EN/IEC60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
EN/IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should 
be at least 30%.
Electrical fast transient/burst
EN/IEC61000-4-4
±2kV for power supply lines
±1kV for input/output
±2kV for power supply lines
±1kV for input/output
Mains power quality should be that of a typical commercial or 
hospital environment.
Surge
EN/IEC61000-4-5
±1kV line to line
±2kV lines to earth
±1kV line to line
±2kV lines to earth
Mains power quality should be that of a typical commercial or 
hospital environment.
Voltage dips, short interruptions and 
voltage variations on power supply input 
lines
EN/IEC61000-4-11
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
Mains power quality should be that of a typical commercial 
or hospital environment. If the user of the VarioSurg requires 
continued operation during power mains interruptions, it 
is recommended that the VarioSurg be powered from an 
uninterruptible power supply or a battery.
Power frequency (50/60Hz)
magnetic field
EN/IEC61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical commercial or 
hospital environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.