9.5 End of life
The REHAB SERIES contains materials that can be
recycled and/or are noxious to the environment.
Specialized companies can dismantle the unit and
sort out these materials. When you dispose of the
unit, find out about local regulations concerning
waste management
Stimulation is
uncomfortable.
Stimulation is
ineffective
Intensity is too high
Electrodes are too
close together
Damaged or worn electrodes
or lead wires
Electrode active area size is
too small
Improper electrode
Unknown
Decrease intensity .
Reposition the electrodes.
Electrodes must be a minimum of 2
inches apart.
Replace electrodes with ones that have
an active area no less than 25.0cm .
Reposition electrode
Contact clinician
Replace