10.1 Ultrasound parameters
Frequency.........................................................................1 MHZ,±10%;3 MHZ,±10%
Duty Cycles.............................................................10%-100%,Continues,stepping 10%
Pulse duration...........................................................1 9 ms ± 10 % (set by duty cycle)
PulseFrequency....................................................................................1648100Hz
Output Power
'XW\IDFWRUIRUFP²........................................................................0.5W-10.0W
'XW\IDFWRUIRUFP²........................................................................0.5W-15.0W
'XW\IDFWRUIRUFP²..........................................................................0.1W-2.0W
'XW\IDFWRUIRUFP²..........................................................................0.1W-2.0W
Output accuracy........................................± 20% (for any level above 10% of maximum)
Peak out Amplitude Duty
'XW\IDFWRU:FPò
'XW\IDFWRU:FPò
Treatment timer.............................................................................0 - 30 min ± 0.1 min
5 cm² Applicator
ERA (Effective Radiation Area) ................................................................................5cm²
Beam type:
1 MHz.........................................................................................................Collimating
3 MHz.........................................................................................................Collimating
BNR (Beam Non-uniformity Ratio)...............................................................5:1 maximum
1cm² Applicator
ERA (Effective Radiation Area).................................................................................1cm²
Beam type:
1 MHz..................................................................................................................1cm²
3 MHz..................................................................................................................1cm²
BNR (Beam Non-uniformity Ratio)...............................................................5:1 maximum
10. Specifications
73
Rehab-Series

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nu-Tek UT2200 and is the answer not in the manual?

Nu-Tek UT2200 Specifications

General IconGeneral
BrandNu-Tek
ModelUT2200
CategoryMedical Equipment
LanguageEnglish

Related product manuals