SafetyȱNotificationsȱȱ
Type / Degree of Protection Classification / Identification / Warnings Symbol
Type of protection against electric shock Class II equipment
The degree of protection against electric shock Type B applied part
The degree of protection against the ingress of liquids Not protected Na
The degree of safety in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide
Not suitable Na
The mode of operation Continuous Na
Information regarding potential electromagnetic or other
interference and advice regarding avoidance
The NuStep
®
TRS 4000 Recumbent Cross Trainer uses
electromagnetic and RF energy only for its internal function. Therefore,
its EMC and RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment..
Na
EMC warnings and tables required by IEC 60601-1-2 See EMC tables below Na
ID of specified optional external power supplies or battery
chargers necessary to ensure compliance with the
requirements of IEC60601
An external power supply is optional due to AA battery operation, but
when required, the following power supplies must be used:
Europe: DPD090050E-P5-TK
Japan: 1J35-090030D-024
US & Canada: DPD090050-P5P-TK
Na
ID of any risks associated with the disposal of waste products,
residues, including disposal of the equipment itself at the end
of its useful life.
The NuStep
®
TRS 4000 Recumbent Cross Trainer contains electronic
circuit assemblies and alkaline batteries that may require specific local
disposal or recycling procedures.
The specification of the environmental conditions of transport
and storage (also marked on the outside of the packaging.
The NuStep
®
TRS 4000 Recumbent Cross Trainer can be:
a) safely transported and stored in these conditions
-20º to 54ºC; 85% non-condensing humidity; 100 Pascals pressure
b) operated in these conditions
5º to 50ºC; 85% non-condensing humidity; 100 Pascals pressure
Na
A description of the means for the isolation of the equipment
from the supply.
The NuStep
®
TRS 4000 Recumbent Cross Trainer can be isolated by
unplugging the power supply from the wall and run on batteries alone.
The power supply has an isolation transformer and fuseable link.
Na
Indication that the equipment is energized. No such indicator provided. Na
Indication of long term battery storage If the NuStep
®
TRS 4000 Recumbent Cross Trainer equipment is not to
be used for longer than 3 months, please remove batteries.
Na
ManufacturerȱandȱEquipmentȱ
Identificationȱ
The serial number is label is located at the top of the electronic tube directly behind the display.
A Manufacturer’s name and address F Quality management system registered to ISO standard by NSF-ISR
B Model number and description of product G International patent protection for the product
C Compliance with directives and standards H Tells you to consult the User’s Manual
D Equipment classification I Type B applied part for electrical safety
E Serial number and date of manufacture J CE mark for Europe
I