1. INTRODUCTION
1.1. USE AND FUNCTION
EN 032, EN 055 and EN 120 incubators are designed to provide excellent incubation conditions
for applications in biology and microbiology laboratories such as medical and veterinary fields;
research and quality control examinations in pharmaceutical, food and cosmetics industries and
biotechnology.
EN 032, EN 055 and EN 120 incubators provide temperature range between the ambient
temperature plus 5C and 100C and keeps temperature stable within given tolerances. The
incubators have timer which can be programmed up to 99.9 hours including hold position and
the incubators stop to operate after the set time is completed. Cross contamination risk is
minimized by means of natural air circulation.
Stable and homogeneous temperature in the chamber is provided by means of the natural air
circulation through the stainless steel heaters placed onto the three outer surface of the
chamber.
The insulation proportional to the dimensions of the incubator not only provides a homogenous
temperature distribution but also makes the operation of incubator economics. The incubator
ensures reliable working conditions by N-Prime
TM
programmable microprocessor with high
control accuracy. Adjustable safety thermostat offers an additional safety for your operations.
Microprocessor control system will shut down the temperature sensor and in case of
malfunctions that may occur in the control system, alarm system will be activated and the user
is warned visually and audibly. The study data are recorded in the memory and can be
transferred to external USB memory. At the same time, unauthorized persons have been
blocked permission to change parameters with improved password menu.
EN 032N, EN 055N and EN 120N Incubators are manufactured according to the following
standards: EN 61010-1, EN 61010-2-010, EN 50419, EN 61000-6-3, DIN 12880.
This device is in compliance with WEEE Regulation.
If the warnings mentioned in this manual are not considered, NUVE will not be responsible
from their results.