5.7. CONNECTION TO THE MAINS, WATER AND AIR SUPPLY
5.7.1. CONNECTION TO THE MAINS
The unit is fed by 3-phased network supply. Please make sure that the supplied mains match
the required power ratings. If not, please provide an extra line to support.
The neutral and grounding cables connecting the socket to the general panelboard shall be
separate lines.
Please make sure that the connection cables between the socket and the panelboard are equal
or greater in cross-section than the power cables of the unit.
The panelboard where the socket is connected shall be fuse protected.
5.7.2. CONNECTION TO WATER SUPPLY
There exist a supply water inlet line and a waste water discharge line on the unit.
The water from the mains shall be connected to the water inlet of the unit after being treated
by a suitable water softening unit. It shall be always kept in mind that the supply water quality
has a direct effect on the lifetime of the equipment.
The waste water discharge line shall be connected to the general drain system. The piping of
the drain system shall be resistant to a continuous water flow at 65C and a temporary (2-3
minutes) water flow at 100C.
A regulator should be used on the lines where the water supply pressure is over 4
bars.
5.7.3. CONNECTION TO AIR SUPPLY
There is air inlet on the top of the sterilizer.
Dry air line should be connected and it should have 5 to 6 bars of pressure.
6. OPERATING PRINCIPLES
6.1. PRIOR TO OPERATION
Turn on the unit by using the On-Off switch on the left side
See that the microprocessor control system is activated.
NUVE logo appears on the screen.
After NUVE logo disappears, Menu page appears.