6.4.6. STACKING STERILIZATION CONTAINERS
When arranging sterilization containers, care
should be taken that drops of condensate do
not wet items being sterilized beneath, but
can flow away to the base of the chamber. The
best arrangement is a stack of sterilization
containers of the same size, so that
condensate can flow down the sides.
Figure 10 - Stacked sterilization containers
7. PERIODIC MAINTENANCE AND CLEANING
7.1. PERIODIC MAINTENANCE
Cleaning and periodical maintenance operations should be performed the manometer
pressure is 0 bar, the lid is open position and the chamber is cold. During cleaning and
maintenance operation, remove the plug from the outlet.
All the safety valves subjected to direct vapor pressure shall be changed once in five years by
the authorized personnel.
After each 2000 runs the instrument should be controlled by authorized technical service
personnel.
The door gasket shall be replaced by the authorized personnel after each 500 runs or once
per six months.
The air filter shall be replaced after each 300 runs.
It is recommended that the Bowie&Dick Test shall be performed weekly, while the Vacuum
Test shall be executed at the beginning of every working day while the chamber is not yet
heated; to assure efficient functioning of the unit.
Check the safety valve by operating it.
Remove the cover of the autoclave, check and tighten the
ports and valves.
Check that the autoclave is leveled.
Check the continuity of the grounding connections.
Check the safety elements (safety valve, safety and cut-off
thermostats and lid locking mechanism).