31
In case the error has occurred during program execution; the unit stops the program and either
releases steam or vents the chamber according to the pressure conditions in the chamber.
The unit then goes to the “Finalizing” step to let the door open.
In such a condition, the message “Please Wait” will be displayed on the screen. The user can open
the door as soon as this message disappears.
An error code means that sterilization is not complete and the load is not sterile. The
sterilization shall be repeated.
In case of below written failures, related error codes are shown on the display.
Error 01: Vacuum Time Exceeded
The system pressure has not dropped to the required value within the prescribed duration.
Error02: Steam Release Time Exceeded
The steam within the chamber has not been released within the prescribed duration following
the sterilization phase.
Error 03: Door Lock Open
Door lock has been released during operation.
Error 05: Sensor Failure 1, 2
One or more of the temperature and pressure sensors which perform the test functions of the
equipment are defected.
Error 06: Heater Failure
The error code is displayed when the thermic relays of the heaters give a signal of heater failure.
If only one group of heaters are defective; the unit continues to function while displaying a
message on the screen e.g. “Failure in 1st. group of heaters. Call Service.”
In case more than one group of heaters are defective; the unit stops functioning and displays a
message on the screen e.g. “Failure in 1st. and 4th. group of heaters’ operation terminated. Call
Service”.
Error 07: Air Detector Failure
This message displays when the volume of air in the chamber is above an acceptable limit.
Error 08: Leakage in System, Vacuum Test Failed!
This message is displayed in case the leakage measured during the vacuum test exceeds the
acceptable limit. This error indicates that there is a leakage in the system that leads ambient air
into the chamber.
Error 09: Power Failure
Displayed when the mains supply has gone off and on during a cycle. As all the air and steam lines
will be closed during the power failure, the steam-if any-in the chamber shall be released