3: Using NeoFox with NeoFox Viewer Software
32 013-20000-009-02-201510
3. To save these settings through the NeoFox Sport, select Save Settings from the O2 Setup Menu.
Note
When the device saves a parameter through a flash write, it must save ALL of its savable settings
at the same time. Therefore, actions which may appear to save specific settings to memory will
actually save the entire set of savable parameters. Those actions include downloading calibration
data, performing a single point reset, or manually prompting a flash write by setting the Flashed
Write parameter to Enabled in the Advanced Settings Panel.
Resetting Saved Settings
You can reset the saved settings to their original state.
Procedure
1. Open the device in the NeoFox Viewer and navigate to the Advanced Settings screen.
2. Set the value of the Flash Reset parameter to Enabled. This value may revert to disabled shortly
thereafter. When the device is reset, the saved values revert to their factory settings.
3. Download your calibrations again after resetting the flash memory.
Note
For devices with firmware version 4.00 or above, the APD Bias Voltage (displayed as DAC-APD
Gain) setting value persists through a flash reset.