4: Using NeoFox Sport
013-20000-009-02-201510 39
Sensor Pressure, then this is the sensor pressure.
Fixed Pressure, then this is the fixed pressure value.
None, then this is 101 325.
Sensor Pressure = The pressure (in kPa).
NeoFox Sport Saved Settings
Most of the parameters that can be set through the NeoFox Sport’s user menus can be saved to the
NeoFox Sport module’s flash memory, which allows those settings to be restored the next time that the
device is powered on. See Saved Settings for a list of settings that can be saved.
To save the flash write settings, navigate to the Sensor Setup | Save Settings menu. Note that this saves
ALL of the savable parameters in their current state. Also, performing a Single Point Reset through the
DPU menu may result in all of the savable parameters being overwritten.
Low Battery Detector
The NeoFox Sport contains a low battery detector that prevents inaccurate readings in low battery
conditions. When the device detects that battery conditions are too low for reliable operation, its tau and
oxygen measurements will be set to 0 or -1. NeoFox disables its measurements shortly before the device