A: Calibrating the Wavelength of the QE Pro
26 891-00000-000-02-201603
6. Double-click on each of the calibration coefficients displayed in the Wavelength Coefficients
pane of the USB Programmer screen and enter the new values acquired in Step 5 of the
Calibrating the Wavelength of the Spectrometer section in this appendix.
7. Repeat Step 6 for all of the new values.
8. Click on the Write Coefficients to Device button to save the information, and then select File|
Exit to exit the QE Pro Reflash Utility software.
The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on the QE
Pro.