10 Troubleshooting
28 / 40 Instruction Manual BIOM® 5 (G/55400/0413/en)
Unfocused image BIOM® 5 is maladjusted
A reduction lens is not being
used
The reduction lens is not com-
patible the microscope objec-
tive
Focus the BIOM® 5 in accor-
dance with the instructions
given in the manual
Use a reduction lens
Check the engraving on the re-
duction lens and change the
lens, if necessary (see sect. 9.1,
page 11
Fundus view is too narrow Too much distance between
the ophthalmoscopy lens and
the eye
Magnification of the
microscope system to high
Carefully reduce the distance
using the microscope fine
adjustment mechanism
Reduce magnification of the
microscope
The eye or the lens reflect strongly The microscope light is on Turn the light off, illuminate
only intraocular
The BIOM® 5 cannot be detached
from the adapter
The BIOM® 5 has not been
swiveled out
Vacuum has developed be-
tween the BIOM® 5 and the
adapter, or there are deposits
on the connecting parts
Swivel out the BIOM® 5 for
detachment
Place the BIOM® 5 and adapter
into an ultrasound bath (for
approx. 5 min)
No function whatsoever when the
combination control unit is actuated
The combination control unit
is not connected to the SDI®
4c
The SDI® 4c is not connected
to the 6V-15V power supply
Power failure or power outlet
is not active
These are not active when the
sockets at the microscope
stand are in use
Establish the connection to the
SDI® 4c
Establish the connection to the
6V-15V power supply
Inform the in-house
electrician
Use the 6V-15V plug
transformer
Use the mechanical adjusting
element or adjusting wheel
Activate the sockets in
accordance with the
instructions for the stand
Contact the microscope
manufacturer for assistance
Error Possible Cause Help