9 BIOM® 5 in Use
16 / 40 Instruction Manual BIOM® 5 (G/55400/0413/en)
9.3 Under Sterile Conditions:
Test the Safety Functions
Attention
Risk of injury in the event of malfunction
Only use the BIOM® 5 after the following functions have been tested
and are in perfect working order.
#
Turn the BIOM® 5 towards the adapter plate. Make
sure that the housing body can be shifted without
resistance.
Slide the safety rod of the lens mount up and down
several times to check that it moves smoothly. The
control marking must be visible all times.
Check that the knob turns easily and shorten the
total length until the adjuster is at the uppermost
position.
Check whether all fastening screws are present
(e.g. screws at the feather key of the safety rod).
Fig. 9-6: Test the Safety Functions
1
2
control
marking
3