12 Cleaning, Sterilization and Maintenance
30 / 40 Instruction Manual BIOM® 5 (G/55400/0413/en)
To clean, disinfect and sterilize the equipment, proceed as described in
the Conditioning Manual
.
Attention
Risk of infection if reconditioning is done incorrectly.
Please heed the separate Conditioning Manual.
Note
The sterilization tray from the company OCULUS Optikgeräte GmbH (Art.
No. 55185) can also be used for cleaning the BIOM® components.
Cleaning and Sterilization Instructions
The BIOM® 5, the adapter and the optics can be cleaned in a dishwasher
with a mild alkaline cleaning agent and can be cleaned and disinfected in
the disinfector (refer to the Conditioning Manual).
Due to the special surface finish required for sterilizability of the BIOM® 5,
structural patterns may become visible. This unavoidable minor blemish
does not in any way adversely affect the function, handling and sterilization
of the unit.
All of the components of the BIOM® 5 listed in sect. 16.1, page 36, the
adapter and the presently available reduction lenses and front lenses,
except for the disposable lens sets, can be steam-autoclaved (134°C).
Fig. 12-1: Specially designed insert for sterilization containers. Loaded with the BIOM® 5c setup