13 Cleaning, disinfection and maintenance
Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01) 41 / 60
Cleaning
Switch off the Easyfield®,
sec. 8.2, page 19
.
Disconnect the power plug.
When cleaning, use a damp cloth and make sure that no liquid enters
the Easyfield®.
Clean the plastic surfaces with the appropriate cleaner.
Clean the painted surfaces with the cleaner for painted surfaces.
Clean the lenses in the perimeter cone with a soft cloth or lens brush.
Use alcohol or a lens cleaner if needed.
13.2 Disinfection
Caution
Risk of electric shock if the Easyfield® is not completely disconnected
from the mains for these jobs.
Turn the Easyfield® off,
sec. 8.2, page 19
.
Pull the power plug before cleaning. When disconnecting electrical
connections, pull on the respective plug and not on the cable itself.
Required materials (recommended) or an equivalent:
Pursept
-A Xpress disinfectant cloths,
Schülke & Mayr GmbH
Robert-Koch-Str. 2
22851 Norderstedt I Deutschland
Telefon: +49 40 52100-0
Telefax: +49 40 52100-318
E-Mail: info@schuelke.com
https://www.schuelke.com/de-de/index.php
Disinfect the forehead and chin rest after each examination; disinfect
the housing as needed.
Disinfect the hand held button after each examination.
Disinfect the
eyepatch after each use.