15 Transport and storage
44 / 60 Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01)
15 Transport and storage
The Easyfield®must be properly dismounted and packed before being
transported or stored.
15.1 Disassembly and packing
Select Patient > New Patient / End.
Exit the Patient Data Management system.
Power down the PC/laptop.
Disconnect the power plug from the power jack,
sec. 8.2, page 19
.
Disconnect the connections to the hand-held button, to the netbook/PC/
laptop.
When disconnecting, pull on the respective plug and not on the
cable itself.
Pack up the Easyfield® in the original packing.
15.2 Transport and Storage Information
Storage
Transport
Ambient temperature range
-10°C to +55°C
Relative humidity, including condensation
10% to 95%
Air pressure range
700 hPa to 1060 hPa
Ambient temperature range
-40°C to +70°C
Relative humidity range from
10% to 95%
Air pressure range
500 hPa to 1060 hPa