side (right next to the white “take photo” button)
When the sharpness indicator is on, an indicated
number in a circle will be shown at the bottom left
hand side of the screen. This shows in real time
how high the sharpness level of that particular
image is. The sharpness indicator values will
change through the turning of the focus knob.
When you reach the highest possible value, the
image has reached an optimal level of sharpness
and a photo can be taken. Since the sharpness
indicator is influenced by the number of cells within
each image, it’s important to reach the highest
value indicated for each image, not based on a
specific value each time. (i.e. one image might
have the highest value on the sharpness indicator
at 2300 and another only 500, use the highest
value for each image).
Switch between Methylene Violet (MB) and Methylene Blue
(MV) analysis
You can choose between Methylene Violet or
Methylene Blue as the staining agent for viability
analysis. This is selectable on the start screen –
you will see a blue or violet pipette icon – select
the one you wish to use. The results of the viability
measurements are also displayed in the
corresponding color.
On/Off Switch for Manual Counting
14