OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 48
11. Repeat steps 1 through 3 above to return to the “TC Offset” screen.
Enter the Offset value from step 10 and press ENTER.
If the Offset value is negative, include a “.” at the end. For
example, if the Offset is calculated as -2.2, enter “2.2.” on the
Model 900.
12. Press NEXT to go to the “TC Slope” screen. Enter the Slope value
from step 10.
13. Press ENTER, then CLEAR/STOP to return to the main screen.
14. The Model 900 should now display the same temperature (± 0.2°C) as
the calibrated thermometer.