What to do if OfiTE Laboratory Equipment will not build pressure?
- BbillysmithAug 20, 2025
If your OfiTE Laboratory Equipment is not building pressure, there could be several reasons. First, ensure the cell cap o-ring isn't leaking and replace it if necessary. Check all fittings for leaks and tighten any loose connections. Other potential causes include a broken pump (not cycling or constantly cycling, in this case it should be replaced), a blown rupture disk (needs replacement), an open pressure release valve (close it completely), or the Fill/Drain Cell switch being set to Drain (switch it to Fill). Also, check the oil level in the reservoir and add oil if needed. Make sure air is properly supplied to the unit, empty any water from the air filter, or consider replacing the air solenoid if it is damaged.