OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 33
Standalone
Mode
Managing Tests
The on-board control panel has eight slots for storing tests. The rst slot is
reserved for a pre-dened API Schedule 5 test. This test cannot be edited.
The API Schedule 5 test has the following parameters:
- Motor Speed: 150 RPM
- Starting Pressure: 1,000 PSI
- Starting Temperature: 80°F
- Final Temperature: 125°F
- Final Pressure: 5160 PSI
- Time to Bottom Hole: 28 Minutes
The next three slots contain pre-dened test that can be edited either through
the on-board controls or through the software.
- Slot 1: API Schedule 5
- Slot 2: Casing / Liner
- Slot 3: Squeeze
- Slot 4: Hesitation Squeeze
The remaining four slots are temporary slots for custom tests. You can either
create a new test or edit an existing test by copying it to a temporary slot
using the “Copy Test to Temp” feature described on page 35.
In order to run or edit a test, you must rst load the test into memory using
the “Select New Test” option.
Selecting a New Test:
1. Push the Scroll Wheel to access the Main Menu.
2. Scroll to option 2, “Select New Test”. Push the Scroll Wheel to enter.
3. Select a test from the list. Push to enter.
Once a test is selected, it can be run, edited, or copied.