EasyManua.ls Logo

OfiTE 2025 - Managing Tests

OfiTE 2025
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 33
Standalone
Mode
Managing Tests
The on-board control panel has eight slots for storing tests. The rst slot is
reserved for a pre-dened API Schedule 5 test. This test cannot be edited.
The API Schedule 5 test has the following parameters:
- Motor Speed: 150 RPM
- Starting Pressure: 1,000 PSI
- Starting Temperature: 80°F
- Final Temperature: 125°F
- Final Pressure: 5160 PSI
- Time to Bottom Hole: 28 Minutes
The next three slots contain pre-dened test that can be edited either through
the on-board controls or through the software.
- Slot 1: API Schedule 5
- Slot 2: Casing / Liner
- Slot 3: Squeeze
- Slot 4: Hesitation Squeeze
The remaining four slots are temporary slots for custom tests. You can either
create a new test or edit an existing test by copying it to a temporary slot
using the “Copy Test to Temp” feature described on page 35.
In order to run or edit a test, you must rst load the test into memory using
the “Select New Test” option.
Selecting a New Test:
1. Push the Scroll Wheel to access the Main Menu.
2. Scroll to option 2, “Select New Test”. Push the Scroll Wheel to enter.
3. Select a test from the list. Push to enter.
Once a test is selected, it can be run, edited, or copied.

Related product manuals