Quick Start
Completing a Test
When the test is complete, the red light on the Status Indicator will turn on
and an alarm will sound. To acknowledge the alarm, push the Scroll Wheel.
1. Turn the HEAT, MOTOR, and DC VOLT switches off.
2. When the cell temperature has reached 150°F, turn off the PUMP switch
and slowly open the Pressure Release valve.
Never release pressure while the temperature is above 200°F
(93.3°C).
3. Turn the CYLINDER switch to DRAIN and turn the AIR TO CYLINDER
switch on. This will allow air into the test cell and force the oil back into
the reservoir.
4. When you hear air venting from the Consistometer, turn off the AIR TO
CYLINDER switch.
5. Remove the thermocouple from the cell cap.
6. Unscrew the cell cap and remove it from the test cell.
7. Remove the potentiometer and the slurry cup from the test cell.
8. Turn the COOL switch off.
9. Return the cell cap to the test cell to prevent dust and other material from
entering the cell.
10. Turn the MAIN switch off.