Calibration
Temperature
The HTHP Consistometer has two thermocouples. One measures the
temperature of the bath, the other measures the temperature of the sample.
Both will require calibration quarterly or if either thermocouple is replaced.
1. Push the Scroll Wheel to access the Main Menu.
2. Scroll to option 5, “Utilities”. Push the Scroll Wheel to enter.
3. To calibrate the sample thermocouple, scroll to option 4, “Calibrate
Sample Temp”. To calibrate the bath thermocouple, scroll to option 5, “Cal
Bath Temp”. Push the Scroll Wheel to enter.
4. Place the thermocouple in a dry block calibrator.
5. Set the calibrator to the value displayed on the screen.
6. Scroll to adjust the offset and make the two temperature readings the
same. Push the Scroll Wheel to save the calibration.