OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 36
Standalone
Mode
Utilities
Utilities Menu
1. Testing Options
2. Calibrate Potentiometer - See page 38.
3. Calibrate Pressure - See page 41.
4. Calibrate Sample Temp - See page 42.
5. Calibrate Bath Temp - See page 42.
6. Error Code History
7. Set Time/Date
Testing Options
1. Temp Units: Deg F / Deg C
2. Display Temp: Bath / Sample - Species whether the display screen
on the front panel will display the reading from the bath thermocouple
or the sample thermocouple. (Note: The Eurotherm temperature
controller will always show the reading specied in “Control Temp”.)
3. Control Temp: Bath / Sample - Specied whether the temperature
controller uses the reading from the bath thermocouple or the sample
thermocouple to control the heaters.
4. Press Units: PSI / MPa (Changing this option will reboot the
Consistometer.)
5. Test RPM - Sets the rotational speed of the slurry cup. (1 – 300 RPM)
6. Cooling OFF Temp: After a test, when the temperature reaches this
point, the cooling will be turned off.
7. Max Bc for Green Light: The consistency at which the yellow light
comes on.
8. Max Bc for Green and Yellow Lights: The consistency at which the
green light turns off.
9. Min Bc for Red Light: The consistency at which the yellow light turns
off and the red light comes on.