Running a Test:
1. Push the Scroll Wheel to access the Main Menu.
2. Scroll to option 1, “Run/Edit Test”. Push the Scroll Wheel to select.
3. Push the Scroll Wheel to run the test.
4. Prepare the slurry cup and test cell for testing. See page 16 for
instructions.
5. Push the Scroll Wheel to start the test.
Editing a Test:
1. Push the Scroll Wheel to access the Main Menu.
2. Scroll to option 1, “Run/Edit Test”. Push the Scroll Wheel to select.
3. Scroll to the parameter you want to change. Push the Scroll Wheel to
select.
4. Scroll to select a new value. Push the Scroll Wheel to enter.
5. Repeat steps 3 and 4 for every parameter you want to change.
The API Schedule 5 test in Slot #1 cannot be edited. However, it can be
copied into a temporary slot.
There are two types of tests: normal and hesitation squeeze. The two types
have different parameters.
Normal:
- Surface Retention Time: Time period before the test begins where the
motor is on but the heater and pressure are off.
- Starting Temp: The initial temperature set point for the test.
- Starting Pressure: Initial pressure for the test.
- Time to Bottom Hole: This is the amount of time (in minutes) the test
will take to fully heat and pressurize the slurry.
- Bottom Hole Temp: This is the nal test temperature.
- Bottom Hole Pressure: This is the nal test pressure.