What to do if OHAUS Scales show Error 8.3?
- NnancyromanAug 31, 2025
If your OHAUS Scales display Error 8.3, it means the weight reading exceeds the overload limit. Reduce the weight on the pan to resolve this issue.
What to do if OHAUS Scales show Error 8.3?
If your OHAUS Scales display Error 8.3, it means the weight reading exceeds the overload limit. Reduce the weight on the pan to resolve this issue.
What does Error 9.5 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 9.5, it indicates that calibration data is not present. Calibrate the scale.
Why is my OHAUS Measuring Instruments unable to calibrate?
If you are unable to calibrate your OHAUS Measuring Instruments, ensure that the LFT security switch is turned off. Also, verify that you are using the correct calibration mass.
What to do if my OHAUS Accessories indicator is unable to turn on?
If your OHAUS Accessories indicator isn't turning on, there are several possible causes. First, ensure the power cord is correctly plugged into both the indicator and a working power outlet. Check the power source to confirm it's supplying electricity. If you're using a battery-powered model, try recharging the battery. If none of these steps work, service may be required.
What does Error 8.92 and Error 8.94 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 8.92 or Error 8.94, it indicates incorrect data from the EasyConnectTM base under LFT ON status. Break the seal or replace the original base/indicator. Service may be required.
What does Error 8.1 mean on my OHAUS Accessories scale and how do I fix it?
Error 8.1 on your OHAUS Accessories scale indicates that the weight reading exceeds the Power On Zero limit. Remove weight from the pan. If the problem persists, recalibrate the scale.
What does error 8.8 mean on OHAUS Scales?
If your OHAUS Scales display Error 8.8, the factory calibration data in the EasyConnectTM module on the load cell cable is not valid under LFT off status. Calibrate the scale to resolve this.
What does Error 8.2 mean on OHAUS Scales?
If your OHAUS Scales display Error 8.2, it means the weight reading is below the Power On Zero limit. To fix this, add weight to the pan. If the issue remains, recalibrate the scale.
What does error 8.4 mean on OHAUS Scales?
If your OHAUS Scales show Error 8.4, the weight reading is below the Underload limit. Add weight to the pan. If the problem continues, recalibrate the scale.
What does Error 8.1 mean on OHAUS Scales?
If your OHAUS Scales display Error 8.1, it indicates that the weight reading exceeds the Power On Zero limit. Remove weight from the pan. If the issue persists, recalibrate the scale.
Model | i-DT33XW |
---|---|
Display | LCD with backlight |
Application | Weighing, Parts Counting, Checkweighing |
Communication | RS232 |
Tare Range | To capacity by subtraction |
Calibration | External Calibration |
Type | Bench Scale |
Power Supply | AC Adapter |
Approval | NTEP approved |
Operation | AC Adapter (Included) |
Construction | Stainless Steel |
Design Features | Water-resistant |
Units | lb, oz, g, kg |
Crucial safety guidelines and warnings for safe operation and handling of the indicator.
Covers connecting the indicator to power sources like AC or battery supplies.
Provides step-by-step instructions for opening the housing and making internal connections.
Guide to entering and using the weighing mode, including application settings.
Explains how to use the counting application, including setting sample size and APW.
Instructions for using the check mode to compare weights against target limits.
Details on accumulating weights, setting totalization methods, and viewing results.
Comprehensive guide to performing various calibration procedures like zero, span, and linearity.
Steps to configure the indicator for legal for trade applications, including calibration.
Common issues, their probable causes, and recommended remedies for the indicator.