EN-28 PR Series Balance
6.3 Sealing
After the Balance has been verified, it must be sealed to prevent undetected access to the legally controlled
settings. Before sealing the device, ensure the Approved Mode setting in the Balance Setup menu has been set to
ON.
If using a paper seal, place seals over the security switch and the bottom housing as shown.
If using a wire seal, pass the sealing wire through the holes in the security screw and the bottom housing
as shown.
Location of security switch Paper Seal Wire Seal
Figure 6-1. Sealing
Push
Button