Before installing the controller, cut off the power supply. 
Also refer to the Particular Specifications chapter on page 39. 
The controller is intended for use in installation areas that meet Class II 
overvoltage requirements and Degree 2 pollution requirements. 
The electrical connection must be: 
■  Performed by a specialist and conform to current regulations in force ; 
■  Compliant with NF C 15-100. 
Verify the current and the grid power supply (the grid power supply must 
correspond to the supply indicated on the controller front plate). The operating 
voltage is configured at the factory. 
The MX 32 does not have an on/off switch. 
Certain power supplies can cause serious or fatal injury. All 
installation and wiring should be performed before turning on the 
power supply. 
Incorrect installation can lead to measurement errors or system 
failure, all instructions in this manual must be followed carefully to 
guarantee proper system operation. 
Ground Connection Protection 
The controller must be connected to a functional ground connection.  
The ground terminal (yellow) (Figure 4, ref. J) is indicated with the following 
symbol:  . 
The cable used must have a minimum diameter of 1.5 mm² and a maximum 
diameter of 2.5 mm². 
Refer to the wiring examples beginning on page 11. 
Power 
The MX 32 must be protected upstream by a differential bipolar circuit 
breaker. The response curve must be type D.  
Differential bipolar circuit 
breaker caliber 
The cable used must have a minimum diameter of 1.5 mm² and a maximum 
diameter of 2.5 mm².  
There are two options for transformers suggested for use as a power supply 
for the MX 32.  
■  Bobbin transformer (standard default). 
■  Toric Transformer (for connecting special sensors such as flame sensors, 
etc.).