EasyManua.ls Logo

Oldham MX 32 - Module Wiring and Grounding

Oldham MX 32
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MX 32
Instruction Manual
Preventative maintenance
Action or result
Illustration
The CAL indicator blinks.
Selecting the channel to calibrate
Press (ENTER) to confirm.
The Channel 1 icon blinks indicating Channel 1
has been selected.
To select Channel 2, press (+).
The Channel 2 icon blinks indicating Channel 2
has been selected.
Zero settings
Confirm the channel to calibrate by pressing
(ENTER).
- Display of the current value (±0).
X0 = zero settings.
- The icon is displayed.
- The yellow indicator of the corresponding
channel blinks (the channel relays are
blocked).
Make sure that the air is clean, if not, using a test
kit, inject air at the level of the detector at a flow
rate of 60 l/h, then wait for stabilization of the
signal.
If the display is a value other than zero, adjust the
zero measurement using the (+) or (-)
buttons.
Sensitivity settings
Inject the calibration gas (60 l/h) at sensor level
and wait for signal stabilization.
Confirm the entry in the sensitivity settings by
pressing (ENTER).
- Display of the current value (±0).
XF = sensitivity settings (blinking icon).
- The icon is displayed.
- The yellow indicator of the corresponding
channel blinks (the channel relays are
blocked).
If necessary (displayed value other than the value
of the calibration gas), adjust the sensitivity by
using the (+) or (-) buttons.

Other manuals for Oldham MX 32

Related product manuals