Chapter 3 Preparation and Inspection
29
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Inspection of the suction valve (MAJ-207)
Do not use the suction valve (MAJ-207) for more than six
procedures.
This suction valve (MAJ-207) is designed for use with
BF-30/40/60/200/240/260/160 and 180 series endoscopes.
1. Confirm that the valve is not deformed or cracked (see Figure 3.5).
Figure 3.5
2. Confirm that the button can be pushed without excessive resistance.
Button
Suction
connector
Arm