Chapter 3 Preparation and Inspection
31
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Inspection of the single-use biopsy valve (MAJ-210)
The single-use biopsy valve (MAJ-210) is provided in a
sterile condition. Do not open the package until you are ready
to use it.
Do not use the instrument after the expiration date displayed
on the sterile package. Doing so may pose an
infection-control risk.
Do not attempt to reuse or sterilize the single-use biopsy
valve (MAJ-210). This could pose an infection-control risk or
cause equipment damage.
Do not use the single-use biopsy valve (MAJ-210) with the
single use aspiration needle (NA-201SX-4022). Please use
the single use adapter biopsy valve (MAJ-1414). And refer to
the instruction manual for the single use aspiration needle
(NA-201SX-4022).
The single-use biopsy valve (MAJ-210) is designed for use
with BF-30/40/60/200/240/260/160/and 180 series
endoscopes.
1. Inspect the sterile package for tears, inadequate sealing, or water damage.
If the sterile package shows any irregularity, the sterile condition of the
instrument may have been compromised. Use a spare instead.
2. Inspect the biopsy valve for cracks or other damage (see Figure 3.7).
Figure 3.7
Slit
Housing