EasyManua.ls Logo

Olympus bx40 - Page 68

Olympus bx40
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BX40F
Ike
C
INSPECTION
STANDARD
CHECK
ITEMS
AND
STANDARDS
Item
Standard
1.|
Fine
focus
+2um
maximum
in
the
+20um
fine
adjustment
range
sensitivity
2.|
Fine
adjustment
No
feeling
of
play
and
tightness
knob
3.|
Coarse
adjustment
No
play.
Focus
adjustable
with
40x.
knob
(The
knob
tension
should
be
previously
adjusted
to
prevent natural
falling
of
the
stage.)
4
as
*
An
image
should
remain
with
40x
when
the
stage
is
a
SS
SSunene
lowered
and
return
back
to
the
lock
position
after
lock
locking.
The
lock
should
be
definitely
released.
5.|
Field
diaphragm
Smooth
with
a
play
of
1/5
or
less
of
the
scale.
*
The
diaphragm
blade
should
not
stop
or
bind.
6.|
Field
diaphragm
The
field
diaphragm
should
be
seen
with
a
100x
size
and
form
objective.
.
Тре
ratio
of
the
longest
side
to
the
shortest
side
should
be
2:1
or
less
in
the
minimum
dia-
phragm
diameter.
7.
|
Condenser
holder
Should
be
smooth.
stroke
+
Chattering
during
stroke should
be
minimized.
8.|
Condenser
holder
*
The
condenser
holder
should
not
butt
against
the
upper limit
slide
glass
at
the
upper
limit
position.
The
field
diaphragm
should
be
focused
with
a
1.4mm
thick
slide
glass.
9.|
Power
supply
unit
*
When
the
power
is
turned
on,
the
bottom
LED
should
light
with
the
light
intensity
control
set
to
the
minimum
position
and
all
LEDs
should
light
with
the
control
set
to
the
maximum
position.
.
The
light
intensity
control
should
move
smoothly
and
the
LED
should
not
flicker.
When
the
preset
switch
is
turned
on,
the
supply
voltage
should
be
switched
to
a
desired
value.
(The
setting
range
is
1.5
~
5.9V
and
set
to
3.7
~
4.3V
at
the
factory.)
10.|
Mounting
of
-
The
observation
tube,
the
stage,
the
condenser,
modules
the
lamp
house
and
the
revolving
nosepiece
can
be
mounted
easily.
b