Item Specification
7. Condenser Type
U-AAC
Achromat-aplanat
U-SC3
Swing-out
BX3-UCD8A
Motorized, achromat-
aplanat, swing-out
N.A. 1.40 0.9-0.1 1.40-0.2
Aperture iris
diaphragm
Graduated scale provided ––
Applicable
objective
powers
10X – 100X 1.25X - 100X 2X - 100X
8. Operating
environment
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104°F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly through
70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations: ±10%
· Pollution degree: 2 (in accordance with IEC60664-1)
· Installation/Overvoltage category: II (in accordance with IEC60664-1)
9. Storage environment · Temperature: Min. -25°C, Max. 65°C
· Humidity: Min. 0%, Max. 90%