56
Before use
Maximum output voltage at effect level 3 ....................................................................... 645 V
p
Maximum output current .................................................................................................. 1.5 A
Rated load ....................................................................................................................... 500 Ω
PlasmaCut
Bipolar cutting in conductive uid for transurethral resection (TURis).
Waveform ........................................................................Continuous sinusoidal, unmodulated
Crest factor
1)
............................................................................................................... 1.3 to 1.8
Range of power levels
2)
............................................................................................1 to 320 W
Effect levels ...................................................................................................................... 1 to 3
Maximum output voltage at effect level 1 ....................................................................... 597 V
p
Maximum output voltage at effect level 2 ....................................................................... 625 V
p
Maximum output voltage at effect level 3 ....................................................................... 700 V
p
Maximum output current ................................................................................................... 12 A
Rated load ......................................................................................................................... 75 Ω
PK LoopCut
Cutting of varying tissue structures, especially broid tissue with advanced bipolar
technology. Intermittent output for fast and controlled cutting.
Waveform ........................................................................Continuous sinusoidal, unmodulated
Crest factor
1)
............................................................................................................... 1.3 to 1.8
Maximum application time.................................................................................................. 60 s
Range of power levels
2)
............................................................................................1 to 200 W
Effect levels ...................................................................................................................... 1 to 3
Maximum output voltage at effect level 1 ....................................................................... 440 V
p
Maximum output voltage at effect level 2 ....................................................................... 520 V
p
Maximum output voltage at effect level 3 ....................................................................... 590 V
p
Maximum output current ..................................................................................................... 3 A
Rated load ....................................................................................................................... 200 Ω
Limited activation time ........................................................................................................ 60 s
PK MorceCut
Cutting of varying tissue structures, especially broid tissue with advanced bipolar
technology. Intermittent output for fast and controlled cutting and enhanced smoke
reduction.
Waveform ........................................................................Continuous sinusoidal, unmodulated
Crest factor
1)
............................................................................................................... 1.3 to 1.8
Maximum application time.................................................................................................. 60 s
Range of power levels
2)
............................................................................................1 to 300 W
Effect levels ...................................................................................................................... 1 to 3
Maximum output voltage at effect level 1 ....................................................................... 500 V
p
Maximum output voltage at effect level 2 ....................................................................... 500 V
p
Maximum output voltage at effect level 3 ....................................................................... 560 V
p
Maximum output current ..................................................................................................... 3 A
Rated load ......................................................................................................................... 75 Ω
Limited activation time ........................................................................................................ 60 s
PK PureCut
Cutting of varying tissue structures using advanced bipolar technology. Fast Spark Monitor
(FSM) ensures smooth and reproducible cutting in varying tissue, e.g. muscle and fat.
Waveform ........................................................................Continuous sinusoidal, unmodulated
Crest factor
1)
............................................................................................................... 1.3 to 1.8
Maximum application time.................................................................................................. 60 s
Range of power levels
2)
............................................................................................1 to 200 W
Effect levels ...................................................................................................................... 1 to 3