EasyManua.ls Logo

Olympus ESG-410 - Technical Data; ESG-410 Specifications; Ambient Conditions and Plug Requirements

Olympus ESG-410
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
99
Technical data
16 Technical data
16.1 Specications for the ESG-410
Size
Width ......................................................................................................................370 ± 2 mm
Height ..................................................................................................................... 241 ± 2 mm
Depth ......................................................................................................................501 ± 2 mm
Depth including connected cables .......................................................................520 ± 10 mm
Weight ........................................................................................................................... 16.4 kg
Weight of packaging ...................................................................................................... 2.48 kg
Power supply
Voltage range ...................................................................... AC 100 to 120 V/AC 220 to 240 V
Frequency ............................................................................................................ 50 Hz/60 Hz
Maximum input power for AC 100 to 120 V ................................................................ 1150 VA
Maximum input power for AC 220 to 240 V ................................................................ 1500 VA
Fuse ............................................................................................................ 2 x 250 V/T 10 AH
Classications
According to IEC 60601-1, protection against electric shock:
Medical electrical equipment ..........................................................................................Class I
Applied parts .............................................................................................................. Type CF
Protection against harmful ingress of water ...................................................................... IP20
Mode of operation ............................................................................Non-continuous operation
According to applicable European Union harmonization legislation .....................................IIb
Output
High frequency functions....................................................................... Monopolar and bipolar
High frequency at rated load .............................................................................344 to 516 kHz
Maximum HF power monopolar .......................................................................300 W at 500 Ω
Maximum HF power bipolar ...............................................................................320 W at 75 Ω
All modes ...................................25% duty cycle (e.g. activated for 10 s, deactivated for 30 s)
16.2 Ambient conditions for operation, storage and transport
Temperature
For operation and storage .................................................................................. +10 to +40 °C
For transport........................................................................................................-25 to +60 °C
Relative humidity
For operation and storage ............................................................ 30 to 85 %, non-condensing
For transport................................................................................. 10 to 85 %, non-condensing
Atmospheric pressure
For operation ................................................................................ 700 to 1060 hPa (≤ 3000 m)
For storage and transport ............................................................. 500 to 1060 hPa (≤ 3000 m)