Use
‒ A corresponding information message is displayed, which provides information on
remedial actions.
To clear the information message from the touchscreen, perform the remedial action that is
displayed and Tap [OK] on the touchscreen. The information message can also be cleared
by re-activating the power output with the hand switch or foot switch. For precise information
on each information message regarding cause and remedy refer to the following table.
Info
code
Message Cause Remedial action
i0763 Incomplete seal cycle
1. Inspect the seal.
2. Conrm no metal object
is grasped. Remove
excess uids.
3. Reactivate instrument.
The electrosurgical
generator detected a short
circuit. The instrument
jaws have contact with
a metal part or there is
excess uid around the
instrument jaws.
1. Release the hand switch or foot
switch.
2. Open the instrument jaws and
inspect the progress of the seal.
3. Remove excess uids.
4. Conrm that no metal objects
(staples, clips) are between the jaws.
5. Re-position the instrument to avoid
metal contact.
6. Repeat the sealing process to
complete the seal.
i0764 Incomplete seal cycle
1. Inspect the seal.
2. Conrm proper tissue
contact. Grasp more
tissue.
3. Reactivate instrument.
The instrument jaws are
not closed completely.
Insufcient amount of
tissue is grasped.
1. Release the hand switch or foot
switch.
2. Open the instrument jaws and
inspect the progress of the seal.
3. If possible, re-grasp more tissue.
4. Close the instrument jaws
completely.
5. Repeat the sealing process to
complete the seal.
i0765 Incomplete seal cycle
1. Inspect the seal.
2. Hold button until
complete seal cycle.
3. Reactivate instrument.
The seal cycle was
interrupted before
completion due to
releasing the hand switch/
foot switch too early.
1. Release the hand switch or foot
switch.
2. Open the instrument jaws and
inspect the progress of the seal.
3. If the seal is incomplete, re-grasp
the tissue (in the same position or
overlapping the rst seal a little).
4. Close the instrument jaws
completely.
5. Repeat the sealing process to
complete the seal.
i0766 Incomplete seal cycle
1. Inspect the seal.
2. Remove excess
uids. Clean and dry
instrument jaws.
3. Reactivate instrument.
No tissue effect. The
energy is absorbed
by excess uid or by
coagulum that has built up
at the instrument jaws.
1. Release the hand switch or foot
switch.
2. Open the instrument jaws and
inspect the progress of the seal.
3. Remove excess uids.
4. Clean and dry the instrument jaws.
5. Re-grasp the tissue.
6. Repeat the sealing process to
complete the seal.