Installation and Connection
2
2.9 When no Mobile Workstation is Used
23
2.9 When no Mobile
Workstation is Used
1
Confirm that the color video printer (OEP-4) is
turned OFF.
2
Connect the power cord of the color video printer
(OEP-4) to its AC mains power outlet first, then to
a wall mains socket outlet.
3
The following ancillary equipment can be
connected directly to wall mains socket outlets (see
Figure 2.8).
4
The following ancillary equipment should be
connected through an isolation transformer (see
Figure 2.8).
EVIS EXERA video system
center
CV-160
EVIS video system center CV-240, CV-140
VISERA video system center OTV-S7V
Video system OTV-SI
3CCD camera control unit OTV-SP1C,
OTV-SP1C-G
Camera control unit OTV-SX2C
EVIS EXERA xenon light
source
CLV-160
EVIS universal light source CLV-U40
Xenon light source CLV-S40, CLV-S30
EUS EXERA endoscopic
ultrasound center
EU-M60
EUS EXERA compact
endoscopic ultrasound center
EU-C60
High definition LCD monitor OEV191H
LCD Monitor OEV191
High definition monitor OEV181H
Color video monitor OEV143, OEV203
Color video printer OEP-4
Monitor Monitor except OEV

Table of Contents