12
SZ2-STU2
4. To tilt the microscope body to the front or rear, loosen the tilting clamping
lever @. (Fig. 19)
Loosen the lever only by half when tilting the microscope body. If
the lever is loosened completely, the microscope body may turn
suddenly and cause danger.
When a photography unit or TV camera is mounted on the micro-
scope body, do not attempt to tilt it.
5. When observing a tall specimen or using an auxiliary objective with long
working distance, loosen the tilting clamping lever @, revolve the mount
shaft ² by 180° so that it points upward, and mount the microscope
body above the horizontal arm. (Fig. 20)
In this case, the stability is deteriorated extremely. Do not mount a
heavy microscope body (SZX) in this way or mount a photographic
unit on the microscope body mounted in this way.
6. To rotate the microscope body horizontally, loosen the clamping
mechanism of the SZX stand arm or SZX microscope body. (Fig. 21)
The microscope body can be rotated by 180°, or by 90° to the both
left and right. However, do not rotate it too much to prevent collision.
When rotating the microscope body, be careful not to loosen the
drop prevention screw below it.
Fig. 19
Fig. 20
Fig. 21
@
@