DISASSEMBLING/ASSEMBLING PROCEDURE
3 Disassembling/Assembling Procedure
3-1 Top cover
Disassembling
1. Remove the screws HCBK4x8SA (9 pcs.)
and CBK4x6SA (1 pc.) which secure the top
cover.
2. Lift and remove the top cover.
CBK4x6SA
HCBK4x8SA
Assembling
1. Mount the top cover on the main body.
2. Secure the top cover to the main body with
the screws HCBK4x8SA (9 pcs.) and
CBK4x6SA (1 pc.).
3-2 Front panel unit
Disassembling
1. Remove the tube and harness from the
UPUH25U.
2. Remove the harness from the power
switch.
3. Remove the front panel unit from the main
body.
Assembling
1. Mount the front panel unit on the main
body.
2. Connect the harness to the power switch.
3. Connect the tube and harness to the
UPUH25U.
Philips screwdriver No. 2