3.9.1 Settings Menus
The settings menu has 3 tabs. In the General tab, you can do the following:
• Select audible indication. The PC will beep every time either of the two
channels goes to high or low alarm conditions.
• COM port auto detect. The program shows the available COM ports for your
selection.
• Selecting the “Show History Viewer”, provides a log of all the events
happening with the application such as high & low alarm events, start &
stop of the application, etc.
• Selecting the “Save to File”, allows you to save the charted data coming from
the recorder into a data file. When you stop recording, the program will ask
if you would like to save the data.
• Select your Engineering units (Up to 5 characters) for the two channels.
• The menu shows the current input type and range. You can change the input
range by selecting other options. Please note that you can change the input
rang within its type. For example, you can change the input from 4-20 mA
to 0-20 mA, or 0-5 V to 0-10V, etc.
Figure 3-6. Settings Menu - General Tab Screen