EasyManua.ls Logo

Omega CTXL-TRH-W-W9 - 4.3 Dual Thermocouple Input Model

Omega CTXL-TRH-W-W9
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
4
4-4
4.3 Dual Thermocouple Input Model
Display Range
Type K: -100 to 1000ºC (-148 to 1832ºF)
Type J: -100 to 700ºC (-148 to 1292ºF)
Type T: -100 to 300ºC (-148 to 572ºF)
Display Accuracy
Type K: ±2ºC (4ºF)
Type J & T: ±1.5ºC (3ºF)
Standard Chart Scales
Type K: -40 to 560ºC (-40 to 1040ºF)
Type J: -40 to 260ºC (-40 to 500ºF)
Type T: -40 to 160ºC (-40 to 320ºF)
Input Connection: Universal connector
Probe Input: Dual K, J, or T programmable from PC
Open Thermocouple Input: Pens will go to the Home position, Display will
show “Prb Err”. Chart motor will continue to run
Out of Range Input: Display flashes
Alarm set point Resolution: 1ºC or ºF
Alarm Deadband: 3ºC or ºF
Factory Default Settings
Thermocouple Input: Dual Type K
Chart Scale, Both Inputs: -40 to 560ºC (-40 to 1040ºF)
High Alarm Set Point
Chan 1: 1040ºF
Chan 2: 1040ºF
Low Alarm Set point
Chan 1: -40ºF
Chan 2: -40ºF
Chart Speed: 7 Days