Specifications
4
4-3
Dimensions: 33.5 H x 27.1 W x 6.7 cm D
(13
3
⁄16 x 10
11
⁄16 x 2
5
⁄8”) You can double
Weight: Approx. 3.2 Kg (7 lbs) including batteries
4.2 Temperature/Humidity Model
Temperature:
Range: -17.7 to 49ºC (0 to 120ºF)
Accuracy: ±1ºC (2ºF) Display, ±1.5°C (3°F) Chart
Display Resolution: 0.1 Degree F or C
Relative Humidity:
Range: 0 to 98% RH
Accuracy (Display): ±3% RH, From 15-90% RH @ room temperature
±5% RH, Below 15 & above 90% RH @ room
temperature
Accuracy (Chart): ±4% RH, From 15-90% RH
±6% RH, Below 15 & above 90% RH
Display Resolution: 1% RH
Distance for Remote Probe: 40 Feet
Temperature/Humidity Sensor: Digital chip
Probe Disconnect: Pens will go to the Home position, Display will
show “Prb Err”. Chart motor will continue to
run
Alarm Set Point Resolution
Temperature: 0.5 ºC or ºF
Humidity: 1% RH
Alarm Deadband: 0.5ºC or ºF – Temperature
3% RH – Humidity
Factory Default Settings:
Chart Scale:
Temperature: 0-120ºF
Humidity: 0-100% RH
High Alarm Set Point:
Temperature: 120ºF
Humidity: 100% RH
Low Alarm Set Point:
Temperature: 0ºF
Humidity: 0% RH
Chart Speed: 7 Days