Clock Battery Backup: 0.33 F Super Capacitor, Holds Clock information
for about 14 days when main power is removed.
White Box Key: Press & hold for 3 seconds to Lock/Unlock
When in Lock mode, all keys are inactive except
for the Power, Light, Mode, Clock, and the
White Box keys
Pens Going to Home Position: At the Power Up
Chart Door is opened
Input Probe is open
Chart is re-scaled
Audible Alarm: Piezo-electric beeper
Alarm Outputs Two separate relay contacts & two voltage outputs
Relay Contacts: 2A @ 30 Vdc
Voltage Output: 100 mA to drive an external relay
Operating Ambient Temperature: 0 to 49ºC (32 to 120ºF)
Operating Relative Humidity: 0 to 98% RH
Power: Four “D” size alkaline Batteries or DC adaptor
Battery Life: 3 Months under normal conditions
Battery Status Indication: Icon on the LCD
AC Adapter 100 to 240 VAC, 50-60 Hz input
Output: 9Vdc @ 1.7A, UL, CE, FCC Marked
Output Plug (Female): Center Positive, Coax 2.0/5.5/10 mm
Output Cable: 1.8 m (6 Ft.) Long
Serial PC Communication: RS232, 2 way, 9600 Baud, 8 Bits Data, 1 Stop Bit,
No Parity
RS232 Adaptor - RJ12 to 9 pin RJ12 Pin # 9 pin D connector Pin #
D - Sub Connector, Female 3 Tx 2 Rx
4 Rx 3 Tx
5 GND 5 GND
Memory: 256K EEPROM, Circular Buffer - Holds 2.8 Chart
revolutions worth of data
PC Communication – See Chapter 3 for more details
Chart Rescale: Rescale one or both pens
Restore Default Factory settings
Data Transfer: Download stored data and save into a data file
Erase stored data from recorder’s memory
Get Data in Real Time: Read current data from channel 1 and 2
Mounting: Key hole slots for wall mounting
Foot cover for Bench top use