FMA 5400A/FMA 5500A
Mass Flow Controllers
omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
Users Guide
Shop online at

Table of Contents

Questions and Answers

  • D
    Dale HaasAug 16, 2025
    How to troubleshoot lack of reading or output on Omega Controller?
    • M
      Mark BrownAug 16, 2025
      If your Omega Controller shows a lack of reading or output, first check the power supply connection. If that's not the issue, the fuse may have blown. Disconnect the transducer from the power supply and remove any shorting conditions or check the polarities; the fuse resets automatically. Also, the filter screen at the inlet or the sensor tube may be obstructed; flush them clean, disassemble to remove any impediments, or replace them. In some cases, the issue may be a PC board defect, which would require a factory replacement. For FMA 5400A/5500A Series Max Flow 10, 50, 100 L/min, the valve adjustment might be incorrect, requiring a readjustment.
  • C
    Carrie BrooksAug 19, 2025
    Why flow reading does not coincide with the setpoint on Omega Controller?
    • S
      Sean MillerAug 19, 2025
      If the Omega Controller's flow reading doesn't match the setpoint, the gas pressure might be inadequate; ensure you apply the appropriate gas pressure. Alternatively, the filter screen at the inlet could be obstructed; try flushing it clean, disassembling it to remove any blockages, or replacing it. A ground loop, where signal and power supply commons differ, could also be the cause.
  • P
    Patricia HarrellAug 21, 2025
    What to do if Omega FMA 5400A has no response to setpoint?
    • C
      Cody BaxterAug 21, 2025
      If the Omega Controller isn't responding to the setpoint, first ensure adequate gas pressure is applied. If the issue persists, check the cables and all connections for any malfunctions, replacing them if necessary. The setpoint might also be too low, specifically less than 2% of full scale; readjust the setpoint or disable the 2% cutoff feature. For FMA 5400A/5500A Series Max Flow 10, 50, 100 L/min, the valve adjustment might be incorrect, requiring a readjustment.
  • Y
    ywashingtonAug 25, 2025
    Why Omega FMA 5400A shows full scale output at “no flow” condition?
    • K
      kristinscottAug 26, 2025
      If your Omega Controller shows a full-scale output at a “no flow” condition or with the valve closed, the sensor may be defective, requiring a return to the factory for replacement. Alternatively, there might be a gas leak; locate and repair it. For FMA 5400A/5500A Series Max valve Flow 10, 50, 100 L/min, the valve adjustment might be incorrect, requiring a readjustment.
  • C
    ccookAug 27, 2025
    How to fix unstable or no zero reading on Omega Controller?
    • H
      Heather JohnsonAug 28, 2025
      If your Omega Controller displays an unstable or no zero reading, there may be a gas leak. Locate and correct the leak. Another potential cause is a defective PC board, which would require a return to the factory for replacement. For FMA 5400A/5500A Series Max Flow 10, 50, 100 L/min, the valve adjustment might be incorrect, requiring a readjustment.
  • R
    rebecca61Aug 30, 2025
    What causes FMA 5400A/5500A valve does not work in open position in Omega Controller?
    • Y
      Yvonne PottsAug 31, 2025
      If the FMA 5400A/5500A valve on your Omega Controller isn't working in the open position, the valve adjustment might be incorrect, so readjust it. Another cause may be a defect in the PC board, requiring a return to the factory for replacement. Also, check the cable and connectors for any malfunctions, replacing them if necessary. The differential pressure could be too high, so decrease the pressure to the correct level, or there might be insufficient inlet pressure, which needs appropriate adjustment.
  • A
    Alec RobertsSep 2, 2025
    Why FMA 5400A/5500A valve does not work in closed position on Omega Controller?
    • S
      Sarah BuchananSep 2, 2025
      If the FMA 5400A/5500A valve on your Omega Controller isn't working in the closed position, the valve adjustment might be incorrect, so readjust it. A defective PC board could also be the cause, requiring a factory replacement. Check the cable and connectors for any issues, replacing them if needed. Additionally, the orifice might be obstructed; disassemble to remove any impediments or return it to the factory.

Summary

Installation and Connections

Gas and Electrical Connections

Guidelines for gas connections, power supply, 15-pin "D" connector pinout, and ESD precautions.

Principle of Operation

Product Specifications

Operating Instructions

Setpoint Control and Valve Operations

Local/remote setpoint control, valve OFF control, and test/purge functions.

Maintenance and Cleaning

Calibration Procedures

Calibrating Flow Controllers (10-1000 L;min)

Detailed steps for zero, span, linearity, and valve adjustments for all models.

Troubleshooting and Technical Assistance

Troubleshooting Common Issues

Checks for common problems, general guide, and valve-specific issues.

Omega FMA 5400A Specifications

General IconGeneral
DisplayDual 4-digit LED
Input Voltage100 to 240 VAC, 50/60 Hz
Control Output4-20 mA
Alarm OutputRelay contact output
Control Output TypeAnalog
CommunicationRS-485