EasyManua.ls Logo

Omega FMA 5400A - Calibrating Flow Controllers (10-1000 L;min)

Omega FMA 5400A
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FIGURE 7-1 FMA 5400A/5500A SERIES MAX. FLOW 10, 50, 100 L/MIN CALIBRATION
POTENTIOMETER AND JUMPER LOCATIONS (BACK OF FMA 5400A/5500A )
7.2 Calibration of FMA 5400A/5500A Series Max.
Flow 10, 50, 100 L/min Mass Flow Controllers
All adjustments in this section are made from the outside of the meter, there is no
need to disassemble any part of the instrument.
FMA 5400A/5500A Mass Flow Controllers may be field recalibrated/checked for
the same range they were originally factory calibrated for. When linearity adjust-
ment is needed, or flow range changes are being made proceed to step 7.2.4.
Flow range changes may require a different Restrictor Flow Element (RFE).
Additionally, a different Solenoid Valve Orifice may also be required (see Table VI).
Consult OMEGA
®
for more information.
7.2.1 Connections and Initial Warm Up
At the 15-pin “D” connector of the FMA 5400A/5500A transducer, connect the
multimeter to output pins [1] and [2] for 0 to 5 VDC (or pins [9] and [14] for 4 to 20
mA) - (see Figure 2-1).
When using a remote setpoint for flow control, the appropriate reference signal
should also be connected to the 15-pin “D” connector at pins [8] and [10] - (see
Figure 2-1). Power up the Mass Flow Controller for at least 30 minutes prior to
commencing the calibration procedure.
19
NR7
LOCAL SET POINT
POTENTIOMETER
R34 ZERO
POTENTIOMETER
R38 50%
R39 75%
R40 100%
R33
SPAN (10 or 25%)
R1 RESPONSE
TIME AJUSTMENT
NJ1 CONTROL
CIRCUT JUMPERS
R52 10 or 25 %

Table of Contents

Related product manuals